Product Description
The NEW T85 Double Beam Spectrophotometer has wavelength range of 190 – 1100nm with either a fixed 2nm bandpass (T85) or with a variable bandpass of o.5, 1.0, 2.0 and 5nm (T85+). The instrument is supplied with an 8 cell motorized cell changer as standard. The T85 and T85+ are full controlled via an embedded touch screen Microsoft PC Tablet pre-loaded with UVWin 6 and with an Operating system of Windows 8 or 10.
With a wavelength range of 190-1100nm and either a fixed 2nm Bandpass (T85) or variable Bandpass 0.5 - 5nm T85+.
You can export data and store all results. The system has FOUR USB connectors for printers or interface to LIMS to Systems
Cell Changer is fully motorised. Large Cell compartment facilutates accesories such as Flow cells and Tablet Dissolution accessories.
Detailed Specification
Optical System | Double Beam |
Spectral Band width | 2nm Fixed Bandpass (T85) and a variable bandpass of 0.5, 1.0, 2.0 and 5nm (T85+) |
Scan Speed | Selectable |
Working Mode | Embedded Touch Screen PC operating in Microsoft Windows 8(TM ) |
Software Support | UV/Win Software v 6.0 (Full GLP version available on request) |
Standard Functionality | Photometric, Quantitative, Spectrum Scanning with 3D data representation and Kinetic measurements. |
Wavelength Range | 190 – 1100nm |
Wavelength Accuracy | ± 0.3nm (Automatic Wavelength Correction) |
Wavelength Reproducibility | 0.2nm |
Stray Light | ≤ 0.05% T (220nm Nal, 340nm NaNo2) |
Photometric mode | Transmission, Absorption and Energy |
Photometric Range | -0.3 – 3.0 Abs |
Photometric Accuracy | ± 0.002Abs (0-0.5A)
± 0.004Abs (0.5 -1.0A) ± 0.3%T (0-100%T) |
Photometric Reproducibility | 0.001Abs (0 – 0.5A)
0.002Abs (0.5 – 1.0A) 0.15%T (0 – 100%T) |
Baseline Flatness | ± 0.0015Abs (190 – 1100nm) |
Baseline Stability | 0.0008Abs/Hr (500nm, 0Abs, 2nm Spectral Bandwidth 2 hr warm up) |
Photometric Noise | ± 0.001Abs (500nm,0Abs, 2nm Spectral Bandwidth) |
Power Supply | Switchable 120 – 230vAC 50/60Hz |